Content: 60 assays (manual) / 600 assays (microplate) / 600 assays (auto-analyser)
Shipping Temperature: Ambient
Storage Temperature: Short term stability: 2-8oC,
Long term stability: See individual component labels
Stability: > 2 years under recommended storage conditions
Analyte: Ethanol
Assay Format: Spectrophotometer, Microplate, Auto-analyser
Detection Method: Absorbance
Wavelength (nm): 340
Signal Response: Increase
Linear Range: 0.25 to 12 µg of ethanol per assay
Limit of Detection: 0.093 mg/L
Reaction Time (min): ~ 5 min
Application examples: Wine, beer, cider, alcoholic fruit juices, spirits, liqueurs, low-alcoholic / non-alcoholic beverages, pickles, fruit and fruit juice, chocolate products, vinegar, jam, bread and bakery products, honey, soy sauce, dairy products, cosmetics, pharmaceuticals and other materials (e.g. biological cultures, samples, etc.).
Method recognition: Methods based on this principle have been accepted by AOAC (AOAC Method 2019.08, First Action), IFU, EBC Method 9.3.1, MEBAK and ASBC Method Beer 4-F
The Ethanol test kit is a simple, reliable and accurate method for the measurement and analysis of ethanol in beverages and foodstuffs.
Note for Content: The number of manual tests per kit can be doubled if all volumes are halved. This can be readily accommodated using the MegaQuantTM Wave Spectrophotometer (D-MQWAVE).
View our full range of alcohol assay kits.
Advantages
Extended cofactors stability. Dissolved cofactors stable for > 1 year at 4oC.
Simple format – aldehyde dehydrogenase supplied as stable suspension
Very competitive price (cost per test)
All reagents stable for > 2 years after preparation
Rapid reaction
Mega-Calc™ software tool is available from our website for hassle-free raw data processing
Standard included
Suitable for manual, microplate and auto-analyser formats
含量:60次测定(手动)/600次测定(微孔板)/600个测定(自动分析仪)
运输温度:环境温度
储存温度:短期稳定性:2-8oC,
长期稳定性:参见单个组件标签
稳定性:在推荐的储存条件下超过2年
分析物:乙醇
测定形式:分光光度计、微孔板、自动分析仪
检测方法:吸光度
波长(nm):340
信号响应:增加
线性范围:每次测定0.25至12µg乙醇
检测限:0.093 mg/L
反应时间(分钟):~5分钟
应用示例:葡萄酒、啤酒、苹果酒、含酒精果汁、烈酒、利口酒、低酒精/非酒精饮料、泡菜、水果和果汁、巧克力制品、醋、果酱、面包和烘焙制品、蜂蜜、酱油、乳制品、化妆品、药品和其他材料(如生物培养物、样品等)。
方法识别:基于该原理的方法已被AOAC(AOAC方法2019.08,第一行动)、IFU、EB接受
Sometimes the addition of the last assay component can cause a small negative absorbance change in the blank samples due to a dilution effect and in such cases it is recommended that the real absorbance values be used in the calculation of results.
The pH of the assay solution after the sample is added should be the same as that of the assay buffer that is supplied with the kit.
Low sample volumes (e.g. 0.1 mL) are not likely to affect the pH of the assay solution and therefore may not require pH adjustment.
Samples above 0.1 mL are more likely to affect the pH of the assay solution and therefore the pH of these samples should be adjusted as described in the data booklet, prior to addition to the assay.
If you suspect that the Megazyme test kit is not performing as expected such that expected results are not obtained please do the following:
Yes. Perchloric acid extraction can be used in conjunction with this kit, and should be performed as follows:
WARNING: If you have not worked with perchloric acid before, you must consult your safety officer for advice. Also, depending on the nature of the samples, it may be possible to reduce the concentration of perchloric acid, to for example 0.3 M (i.e. in the case of plasma). It is thus very important to determine if this is possible for each type of sample used, in order to reduce the risk from working with concentrated perchloric acid.
Liquid samples:
Solid samples:
This can be a common observation when using K-ETOH, however, this problem is very rarely caused by faulty kit components and most commonly caused by equipment or environmental factors. K-ETOH is an extremely sensitive assay so extremely small amounts of ethanol present on any equipment that comes into contact with any of the assay reagents or ethanol absorbed from the atmosphere will affect the K-ETOH assay results. Ethanol kits from all manufacturers are extremely sensitive, and thus certain considerations must be taken into account, especially by certain users, such as in wineries or other laboratories where large quantities of ethanol (or other primary alcohols) are present. However, if the following precautions are observed, accurate results can be achieved with ease:
1. Water Quality - the same water (as ethanol / alcohol free as possible of course) should be used for both the assays and sample dilutions. The best method is to test your source of deionised / distilled water carefully before the analysis session. For example, let the machine producing the water operate for a minute or so, and then test it. When the quality is acceptable [in terms of a very small (i.e. < 0.100) absorbance rise in the blank reaction], put approximately 2 litres or so (in one container, for example a 2 L Duran bottle) aside for the entire analysis session. It is important this bottle (along with all other reagents in fact) be kept airtight, to avoid ethanol being picked up from the air. This occurs very quickly, even in laboratories where ethanol is not used, so in a winery for example, such interference is very rapid and significant.
Contaminated Pipettes - great care should be taken not to use pipettes that were previously employed to perform dilutions (or other manipulations) of wine / concentrated ethanol / alcohol standards, for the assays themselves. The ethanol will go into the vapour phase inside the pipette, and then back into the assay. Thus, one step dilution of samples such as wine is very advisable (i.e. 0.1 mL in 100 mL of water for a 1 in a 1000 dilution).
2. Plastic-ware Contamination - certain pipette tips and cuvettes contain trace amounts of alcohol (i.e. a higher primary alcohol, not ethanol itself) that can lead to elevated / inconsistent results. If this is suspected, all plastic-ware should be washed with deionised / distilled water and dried properly before the experiment. This removes any traces of primary alcohols from the plastic-ware (NOTE: all short-chain primary alcohols react in the system).
3. Blank Reaction - as the reagents themselves will continually pick ethanol / alcohols up from the atmosphere (regardless of the precautions taken), the blank reaction will tend to yield a higher final absorbance value (relative to water or air), as the analytical session progresses (or in the long-term as the age of the kit increases). It is thus very important to run a blank with every set of samples, rather than just one at the beginning or end of the analytical session. Also, as there is a greater chance of sporadic interference with this kit, it is highly recommended that two blank reactions are run, so as to avoid a single faulty blank reading affecting all the results.
4. Environment – as a spectrophotometer is a central piece of equipment in a laboratory, there is little that can be done regarding the location (see (ii) & (iii) below). However, attempting to perform ethanol analyses where a large number of open wine samples, for example, are being manipulated / stored will cause interference. The reasons are related to the very large concentration of this analyte in these samples (i.e. ~ 100 mg/mL), the volatile nature of ethanol, and the increased sensitivity of the assay, as compared to other enzymatic test kits (i.e. two NADH molecules are produced for every ethanol molecule in the sample). The solution is obviously to minimise the potential for interference as much as possible, but this will be very situational:
The final pH of the kit assay after the sample is added should not change from what it should be (as stated in the kit for the assay buffer). If it does change then the sample will require pH adjustment. In most cases the sample volume being used is low relative to the final assay volume and in this case the pH of the kit assay is unlikely to be affected.
The kit assay may work for biological fluids assuming that analyte is present above the limit of detection for the kit after any sample preparation (if required). Centrifugation of the samples and use of the supernatant directly in the kit assay (with appropriate dilution in distilled water) may be sufficient. However, if required a more stringent sample preparation method may be required and examples are provided at the following link:http://www.megazyme.com/docs/analytical-applications-downloads/biological_samples_111109.pdf?sfvrsn=2
The test kit has not been tested using biological fluids as samples because it is not marketed or registered as a medical device. This will therefore require your own validation.
Where the amount of analyte in a liquid sample is unknown, it is recommended that a range of sample dilutions are prepared with the aim of obtaining an absorbance change in the assay that is within the linear range.
Where solid samples are analysed, the weight of sample per volume of water used for sample extraction/preparation can be altered to suit, as can the dilution of the extracted sample prior to the addition of the assay, as per liquid samples.
For users who are not familiar with how to use the Megazyme tests kits then it is recommended that they follow this example, e.g. D-Fructose/D-Glucose Assay kit K-FRUGL (http://secure.megazyme.com/D-Fructose-D-Glucose-Assay-Kit):
1. The kit components are listed on pages 2-3 of the kit booklet.
2. Prepare the kit reagents as described on page 3.
3. For separate measurements of glucose and fructose follow procedure A on page 4.
4. Pipette the volumes listed for water, sample, solution 1 and solution 2 into 3 mL, 1 cm pathlength cuvettes. Duplicate sample assays and duplicate blanks are recommended. Mix the contents of each cuvette by inversion (seal the cuvette using parafilm or a plastic cuvette cap – do not use a finger) then after ~3 min record the first absorbance reading of each cuvette at 340 nm (this is reading A1).
5. Then add suspension 3 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then record the absorbance reading of each cuvette at 340 nm (this is reading A2). NB. It is essential that the reaction is compete. To assess this, record the absorbances at ~ 2 minute intervals and until the absorbance plateaus. A stable absorbance indicates that the reaction is complete. If the absorbance continues to increase then continue to record absorbances until it plateaus and only then record absorbance reading A2.
6. Then add suspension 4 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then take absorbance reading of each cuvette at 340 nm (this is reading A3). NB. As above, assess that the reaction has completed by take subsequent readings at ~2 min intervals.
7. For simple, automated results analysis, input the absorbance readings (A1, A2, A3) for samples and blanks into the K-FRUGL MegaCalc.
To ensure that the assay is working, and being performed correctly it is recommend that the test is performed using the standard sample that is provided with the kit and to obtain the expected values before proceeding to test real samples.
It is recommend that new users also watch this video which highlights how to perform the assays.
Many of the other Megazyme test kits follow a similar format.
The volume/weight of sample and total volume of the extract can be modified to suit the sample. This will ultimately be dictated by the amount of analyte of interest in the sample and may require empirical determination. For low levels of analyte the sample:extract volume ratio can be increased (i.e. increase the sample and/or decrease the total extraction volume).
Alternatively, for samples with low concentrations of analyte, a larger sample volume can be added to the kit assay. When altering the sample volume adjust the distilled water volume added to the assay accordingly so that the total assay volume is not altered.
For samples with low concentrations of analyte the sample volume used in the kit assay can be increased to increase sensitivity. When doing this the water volume is adjusted to retain the same final assay volume. This is critical for the manual assay format because the assay volume and sample volume are used in the calculation of results.
The majority of the Megazyme test kits are developed to work in cuvettes using the manual assay format, however the assay can be converted for use in a 96-well microplate format. To do this the assay volumes for the manual cuvette format are reduced by 10-fold. The calculation of results for the manual assay format uses a 1 cm path-length, however the path-length in the microplate is not 1 cm and therefore the MegaCalc spreadsheet or the calculation provided in the kit booklet for the manual format cannot be used for the micropalate format unless the microplate reader being used can.
There a 3 main methods for calculation of results using the microplate format: